RESEARCH PROTOTYPE
Femoral depot.
A synthetic recess in a CT-derived left femur defines a reference envelope for an idealized tracer depot.
Download reference model 3MF · mm
What we tested.
- Source and defect
- Source specimen s0370. An ellipsoidal cut removed 3.02 cm³. Its original shape provides the known reference for this experiment.
- Bench model
- The reference download is a closed surface with 268,544 triangles in millimetres. After smoothing, the 95th-percentile sampled surface distance from the original reference was 0.371 mm.
- Reconstruction experiment
- The generated surface missed the entire removed region. The interactive surface shows the known removed anatomy.
- Bench simulation
- The blue reference underwent a two-resolution bench test with nominal homogeneous unfilled PEEK properties. One end band was held fixed and the other carried a 1 N axial pull. The fine solve used 43,814 elements and gave a maximum displacement of 0.000112 mm. Two conduction solves held one end at 37 °C and the other at 20 or 45 °C, with insulated free surfaces. The field views preserve the surface used by the solver. Downloads contain the smoothed bench geometry.
- Next experiment
- Test anatomy-specific training and compare reconstructions on independent specimens before evaluating candidate mechanics.
Tracer release through a window
We compared three circular openings on the reference depot. A passive tracer diffused at 0.36 mm²/h toward a perfect sink at the window. Every other surface was sealed.
| Window radius | Released at 24 h |
|---|---|
| 2 mm | 2.59% |
| 4 mm | 8.21% |
| 6 mm | 19.59% |
The 4 mm window came closest to the illustrative 10% release target. This bench calculation uses an inert tracer with assumed diffusivity.
Threshold comparison
Each threshold produced a separate model run. We compared the full output and its largest connected component against the known reference.
| Threshold | Surface | Dice overlap |
|---|---|---|
| 0.50 | Full output | 0.000 |
| 0.50 | Largest component | 0.000 |
| 0.25 | Full output | 0.000 |
| 0.25 | Largest component | 0.000 |
| 0.75 | Full output | 0.000 |
| 0.75 | Largest component | 0.000 |
Wasserthal et al. · University Hospital Basel · Annotated CT dataset v2.0.1 · CC BY 4.0 Adaptations include digital defect creation, surface extraction and field mapping. Annotated CT dataset v2.0.1 · CC BY 4.0